LT30 - Magnescale Co., Ltd.

株式会社マグネスケール

Digital Gauge

LT30

Product TOP

Display unit compatible with our compact, high-precision DK800S series of digital gauges

  • Maxmum display resolution : 0.1μm
  • Zero point detection
  • BCD and RS-232C I/O models are available.
  • Compact and lightweight: DIN standard size (W 72mm x H 72mm)
  • Comparator function
  • Reset/Preset
  • Alarm for exceeded max response speed and disconnected measuring unit
  • Setting value storage
  • 2-axis ADD/SUB (2-axis model only)

Specifications

Specifications

*1 I/O connector Input : Reset, peak-hold start, peak-hold pause, RS trigger (RS-232C models only) Output : Result evaluation (photocoupler)

*2 BCD (36 pin half-pitch connector) Input : Reset, peak-hold start, comparator value selection (4 settings) Output : five digits (open collector)One of current value/maximum value/minimum value/peak-to-peakvalue selected and output. Alerm output

*3 RS-232C (8 pin mini-DIN connector) Reset, preset value setting/recall, peak-hold start, peak-hold pause

*4 RS-TRG pin Trigger input for RS-232C data output

*5 With measuring unit connected.

Magnescale reserves the right to change product specifications without prior notice.

Model LT30-1G LT30-1GB LT30-1GC LT30-2G LT30-2GB LT30-2GC
Display 6 digit, backlight LCD, mode display
I/O Measuring unit input 1 CH 2 CH
I/O connectors*1
BCD*2
RS-232C*3
RS-TRG*4
Reset function Reset key or external input (I/O connectors)
RS-232C command RS-232C command
Preset function Preset value set with preset key, recalled with reset key.
Set or recalled with RS-232C command Set or recalled with RS-232C command
Comparetor function Three-level comparator: Comparator value set with keys on the front panel.
Result evaluation: LED and I/O connector output (photocoupler)
Up to 4 values can be set for comparator (key input). Switched with BCD terminal. Set with RS-232C command Up to 4 values can be set for comparator (key input). Switched with BCD terminal. Set with RS-232C command
Peak hold function Maximum, minimum, and peak-to-peak values. Measuring started by the start input of the I/O connector; update stop by pause input.
RS-232C can set or start. RS-232C can set or start.
Input resolution Selectable from 0.0001 mm, 0.0005 mm, 0.001 mm, 0.005 mm, 0.01 mm
Display resolution Selectable from 0.0001 mm, 0.0005 mm, 0.001 mm, 0.005 mm, 0.01 mm
Direction Can be switched
Reference point function Function use enabled/disabled can be selected (if use is enabled, the unit enters reference point signal input wait status at the same time as power-on).
Maximum response speed 20 MHz (A/B phase difference)
Addition and subtraction function A+B, A-B, B-A can be set with the direction setting.
Alarm Speed over or measuring unit cable disconnected (Displayed on LCD or the I/O connector’s comparator outputs are all “H” (OFF).)
BCD alarm terminal “H” (OFF) BCD alarm terminal “H” (OFF)
Data storage Resolution, direction, comparator value, preset value, modes, etc.
BCD sign, etc BCD sign, etc
key lock function key lock and release.
Temperature Operating:0~40℃, Storage:-10~50℃
Power consumption*5 5 W 5.5 W 5 W 8.5 W 9 W 8.5 W
Mass Approx. 200g Approx. 230g Approx. 220g Approx. 210g Approx. 270g Approx. 230g
Power supply Power input connector (3 pins) : DC9.0-26.4 V
Compatible measuring unit DKseries

Dimensions

Product name 2D 3D
LT30 1G Download
LT30 1GB Download
LT30 1GC Download
LT30 2G Download
LT30 2GB Download
LT30 2GC Download

*Please contact our sales department.

Instruction manual

Product name Japanese English Deutsch Simplified Chinese
LT30 Download Download Download Download
LT30 Supplement Download Download Download

*Please contact us if you cannot download the file.

Accessory

LT30

Output cable

DZ252 RS-232 cable D-Sub 9-Pin female (For LT Series, MG Series)
DZ253A RS-232 cable D-Sub 25-Pin male (For LT Series, MG Series)
DZ254 RS-232C cable flying leads (For LT Series, MG Series)

Catalog

LT30 Series Japanese LT30 Catalog Japanese PDF
LT30 Series English LT30 Catalog English PDF

*Please contact us if you cannot download the file.