Compact, lightweight, and easy-to-mount display unit for DT512

  • Compact size : DIN standard (72 mm x 72mm / 2.83″ x 2.83″ W x H)
  • Resolution: Selectable from 1, 5, 10 μm
  • Suitable for panel mounting
  • Direct interfacing from display unit to PLC or computer
  • Current values, maximum, minimum, peak-to-peak values and GO/NO GO evaluation included as standard functions.
  • ADD/SUB function (2-channel model)
  • Full lineup for various applications

Specifications

Specifications

*1 I/O connector Input : Reset, peak-hold start, peak-hold pause, RS trigger (RS-232C models only) Output : Result evaluation (photocoupler)

*2 BCD (36 pin half-pitch connector) Input : Reset, peak-hold start, comparator value selection (4 settings) Output : five digits (open collector)One of current value/maximum value/minimum value/peak-to-peak value selected and output. Alarm output

*3 RS-232C (8 pin mini-DIN connector) Reset, preset value setting/recall, peak-hold start, peak-hold pause

*4 RS-TRG pin Trigger input for RS-232C data output

*5 With measuring unit connected.

Magnescale reserves the right to change product specifications without prior notice.

Model LT11A-101 LT11A-101B LT11A-101C LT11A-201 LT11A-201B LT11A-201C
Display 6 digit, backlight LCD, mode display
I/O Measuring unit input 1 CH 2 CH
I/O connectors*1
BCD*2
RS-232C*3
RS-TRG*4
Reset function Reset key or external input (I/O connectors)
BCD terminal RS-232C command BCD terminal RS-232C command
Preset function Preset value set with preset key, recalled with reset key.
Recalled with BCD reset terminal Set or recalled with RS-232C command Recalled with BCD reset terminal Set or recalled with RS-232C command
Comparetor function Three-level comparator: Comparator value set with keys on the front panel.
Result evaluation: LED and I/O connector output (photocoupler)
Up to 4 values can be set for comparator (key input). Switched with BCD terminal. Set with RS-232C command Up to 4 values can be set for comparator (key input). Switched with BCD terminal. Set with RS-232C command
Peak Hold function Maximum, minimum, and peak-to-peak values. Measuring started by the start input of the I/O connector; update stop by pause input.
RS-232C can set or start. RS-232C can set or start.
Resolution Selectable from 0.001 mm, 0.005 mm, 0.01 mm
Direction Can be switched
Reference point function Function use enabled/disabled can be selected (if use is enabled, the unit enters reference point signal input wait status at the same time as power-on).
Maximum response speed 100 m/min 100 m/min
Addition and subtraction function A+B, A-B, B-A can be set with the direction setting.
Alarm Speed over or measuring unit cable disconnected (Displayed on LCD or the I/O connector’s comparator outputs are all “H” (OFF).)
BCD alarm terminal “H” (OFF) BCD alarm terminal “H” (OFF)
Data storage Resolution, direction, comparator value, preset value, modes, etc.
BCD sign Data signaling rate, etc. BCD sign Data signaling rate, etc.
key lock function key lock and release.
Temperature Operating:0~40℃, Storage:-10~50℃
Power consumption*5 1.8W 2.9W 2.0W 2.3W 4.0W 2.5W
Mass Approx. 200g Approx. 230g Approx. 220g Approx. 210g Approx. 270g Approx. 230g
Power supply Power input connector (3 pins) : DC9.0-26.4 V
Compatible measuring unit DT512P/N

Dimensions

Product name 2D 3D
LT11A 101 Download
LT11A 101B Download
LT11A 101C Download
LT11A 201 Download
LT11A 201B Download
LT11A 201C Download

*Please contact our sales department.

Manual

Product name Japanese English Deutsch Simplified Chinese
LT11A Download Download Download Download
LT11A Supplement Download Download Download

*Please contact us if you cannot download the file.

Accessory

LT11A

Output cable

DZ252 RS-232 cable D-Sub 9-Pin female (For LT Series, MG Series)
DZ253A RS-232 cable D-Sub 25-Pin male (For LT Series, MG Series)
DZ254 RS-232C cable flying leads(For LT Series, MG Series)