| Model |
DG805 BLM/BLE FLM/FLE
|
DG810 BM/BE FM/FE
|
DG810 BLM/BLE FLM/FLE
|
DG10 BPM/BPE
|
DG10 BSM/BSE
|
DG10 BNM/BNE
|
DG25 BPM/BPE
|
DG25 BSM/BSE
|
DG25 BNM/BNE
|
DG25 BLM/BLE
|
| Measuring length L mm (inch) |
5.2mm
|
10 mm
|
25 mm
|
| Resolution |
0.5/ (1)/ (2)/ (5)/ (10) µm (Switch-selectable)
|
| Accuracy (at 20 °C /68 °F) |
±1 µm
|
| Measuring force(at 20 °C /68 °F) |
Upward |
0.35 ± 0.25N
|
0.4 ± 0.3N
|
All directions 4.9 N or less
|
0.3 ± 0.25N
|
All directions 4.9 N or less
|
0.4 ± 0.3N
|
| Lateral |
0.40 ± 0.25N
|
0.5 ± 0.3N
|
0.6 ± 0.3N
|
0.7 ± 0.35N
|
| Downward |
0.45 ± 0.25N
|
0.6 ± 0.3N
|
0.8 ± 0.35N
|
1 ± 0.4N
|
| Protection grade |
-
|
IP64 Equivalent
|
-
|
IP64 Equivalent
|
-
|
| Mounting stem diameter |
ø 8 (0/ -0.009 mm)/ ø 0.31"(+0/ -0.0004")
L type: ø 9.5(+0/-0.009 mm)/ ø 0.37"(+0/ -0.0004")
|
ø 20(0/ -0.013 mm)/ ø 0.79"(+0/ -0.0005")
|
| Lift lever |
N/A
|
N/A
|
DZ161 (Option)
|
N/A
|
N/A
|
| Air lifter |
DZ801 (Option) Note 1
|
N/A
|
DZ801 (Option) Note 1
|
DZ174-010 (Option) Note 1
|
DZ174-025 (Option) Note 1
|
| Cable length |
5 m/ 16.4"
|
| Operating temperature |
0 to 50 °C / 32 °F to 122 °F
|
| Storage temperature |
20 to 60 °C / 4 °F to 140 °F
|
| Feeler |
Feeler ø 3mm carbide ball with M2.5 x 0.45
screw on fitting end (DZ123)
|
Feeler ø 2.5 mm carbide ball with M2.5 x 0.45 screw on fitting end (DZ122)
|
| Mass (excl. cable unit) |
20g
|
40g
|
100g
|
200g
|
| Model |
DG50 BPM/BPE
|
DG50 BNM/BNE
|
DG100 BPM/BPE
|
DG100 DNM/DNE
|
DG110 BM/BE
|
DG155 BPM/BPE
|
DG205 BPM/BPE
|
| Measuring length L mm (inch) |
50 mm
|
100 mm
|
110 mm
|
155 mm
|
205 mm
|
| Resolution |
0.5/ (1)/ (2)/ (5)/ (10) µm (Switch-selectable)
|
| Accuracy (at 20 °C /68 °F) |
± 1 µm
|
± 2 µm
|
± 2.5 µm
|
± 3 µm
|
| Measuring force (at 20 °C /68 °F) |
Upward |
All directions 6.2 N or less
|
-
|
All directions 9.3N or less
|
-
|
-
|
Resistance against slip: 2.7N (Do not have an internal spring)
|
| Lateral |
0.9 ± 0.4N
|
1.8 ± 0.65N
|
-
|
| Downward |
1.3 ± 0.5N
|
2.7 ± 0.55N
|
1.55 ± 0.15N (Note2)
|
| Protection grade |
IP64 equivalent
|
-
|
IP64 equivalent
|
-
|
-
|
IP64 equivalent
|
| Mounting stem diameter |
ø 20 (+0/-0.013 mm) / ø 0.79" (+0/ -0.0005")
|
ø 20 (+0/-0.03 mm)
ø 0.79" (+0/-0.001")
|
ø 32 (+0/ -0.05 mm)
ø 1.26" (+0/ -0.002")
|
| Lift lever |
N/A
|
DZ161 (Option)
|
N/A
|
DZ161 (Option)
|
DZ161 (Included)
|
DZ161 (Option)
|
| Air lifter |
DZ174-050 (Option) Note 1
|
N/A
|
N/A
|
| Cable length |
5m/ 16.4′
|
2 m/ 6.56′
|
5 m/ 16.4′
|
| Operating temperature |
0 to 50 °C / 32 °F to 122 °F
|
| Storage temperature |
-20 to 60 °C / -4 °F to 140 °F
|
| Feeler |
Feeler tipped with ø 2.5 mm carbide ball with
M2.5 x 0.45 screw on fitting end (DZ122)
|
Feeler tipped with ø 2.5 mm carbide ball with M2.5 x 0.45 screw on fitting end (DZ121)
|
-
|
| Magnetic feeler |
N/A
|
DZ181 (Included) Attraction force : 10N
|
| Mass (excl. cable unit) |
350 g
|
600 g
|
1150 g
|
1100 g
|
1300 g
|
| Model |
DG-M
|
DG-E
|
| Power supply voltage |
DC 5 V ± 5 % (Corresspond to
remote sensing function)
|
DC 11 V to 28 V (Corresspond to remote sensing function)
|
| Power consumption |
300 mA (Maximum)
|
150 mA (Maximum)
|
| Output interface |
Based on EIA-422 (Line driver:AM 26C31 or equivalent)
|
| Output signal |
A/B quadrature signal
|
| Resolution |
0.5,1,2,5,10 µm (0.000025",0.0005",0.0001",0.00025",0.0005") selectable
|
| Maximum response speed |
1 m/s
|
| Minimum phase difference |
200 ns
The output phase difference varies with the measuring speed.
|
| Alarn |
A/B quadrature output of this probe turns to high impedance
for 10 to 20 ms when excessive speed or cable disconnection alarm occurs.
If the cause is removed, the alarm signal is automatically reset.
|
| Output connector |
MDR10214-52A2JL (3M)
|
| Dimensions |
125 x 28 x 28 (mm) (not including projecting parts)
|
| Operating temperature |
0 to 40/ 32 °F to 104 °F
|
| Storage temperature |
-20 to 60/ -4 °F to 140 °F
|
| Mass (excl. cable and gauge part) |
150 g
|